Stat Fax 3300
Biochemistry Semi-Automated Analyzer
Programmable open system
​
The compact design, large LCD screen and user friendly interface make the Stat Fax® 3300 biochemistry analyzer an excellent choice for a variety of laboratory applications.
-
Real-time kinetic plots.
-
Levey-Jennings graphs for QC tracking.
-
Stores calibration and standard curves.
-
Stores and reports patient data.
​
​
Famous For Durability
-
Accepts both round and square test tubes or flow cell.
-
Bichromatic optical system with 6 filters and space for 2 additional filters (340 nm - 700 nm).
-
Calibration traceable to NIST standards
-
Multiple calculation modes: factor, single standard, point-to-point, regression, and index calculation.
-
Reads, calculates, and prints results, 3 seconds per tube.
-
Store up to 120 tests in memory.
-
Stores over 500 patient results and 500 control results.
-
Optional incubation block at 37Ëš and external keyboard.
PRODUCT CHARACTERISTICS
Photometer
-
Tube or cuvette 0 to 2.5 absorbance units.
-
With flowcell 0 to 3.5 absorbance units.
-
Stability Drift or no more than 0.005A in 8 hours/bichromatic.
-
Light source Tungsten halogen lamp, with lamp saver feature.
-
Standard wavelengths 340, 405, 505, 545, 580, 630 nm plus 2 optional.
-
Optional filters 340 - 750 nm.
-
Filter types Ultra long life multiple cavity IAD hard coat.
​​
Sampling
-
Sample volume - flowcell Programmable 250 μL - 750 μL.
-
Flowcell body Surgical grade stainless steel.
-
Minimum Volume - tube 12 mm 1ml.
-
Multiple Calculation Modes.
-
Factor, Single Standard, Point-to-Point, Regression, Index Calculation.
-
Rate (factor/standard) Live graphic display of reaction.
-
Curve fit Linear, point to point, or regression.
-
Miscellaneous blank and curve recall, sample blanking.
Special Features
-
Password protection for test results and parameters.
-
Levey-Jennings QC tracking, patient reports, job creation.
Other Characteristics
-
User tests More than 100.
-
Stored patient results More than 500.
-
Stored control results More than 500.
-
Optional external heat block 18 wells, controlled temp @ 37ËšC.
-
Optional read cell cooling Peltier cooling system - cell only.
-
Cell temperature selections 25ËšC, 30ËšC, or 37ËšC.
-
Dimensions 16” (40 cm) x 14.5” (37 cm) x 5.5” (14 cm) 12” (30 cm) with screen up.
-
Weight (17 lbs) 6.4 kg.
-
Power 100-240 VAC (auto-sensing) 50/60 HZ input, 90 watts.